Triglycerides

Reference

Reference
Nunes LA, Brenzikofer R and Vaz de Macedo D, 2010, Eur J Appl Physiol, 110, 191-8 , Reference change values of blood analytes from physically active subjects

Originally from this reference: Eur J Appl Physiol, Nunes LA, Brenzikofer R and Vaz de Macedo D, 2010


ID: 0268

Scoring

Total Score

B6,7

Parameter Score
1. Scale A
2. Study Population A
3. Samples A
4. Measurand A
5. Pre-Analytical A
6. Replicate Analysis B
7. Steady State B
8. Outlier Analysis A
9. Normality A
10. Variance Homogeneity A
11. ANOVA A
12. CI A
13. No of Results A
14. Concentrations A
15. Total Score B6,7
Comment ATLETES

Item Details

Parameter Score
Matrix Serum
Measurand Triglycerides
Unit mmol / L
Overall Analytical Principle Method Enzymatic analysis - Not specified
Number of Subjects 56
Number of Subjects included in BV estimation 56
Ethinicity Caucasian
Age Mean / Median
Age Min
Age Max
Age SD
Number of Males 56
Number of Females 0
State of Well Being Healthy (excl pregnant)
Study Duration Length 4
Study Duration Units Month
Sampling Intervals, e.g. X sample(s) per Interval Unit 1
Interval Units Month
Sampling Times Start Time: 08:00
End Time:
Number of Samples 4
Average Number of Samples Used for BV Estimation 4.0
Average Number of Replicates 1.0
Estimates of CVi 24.0
Calculated Confidence Intervals of CVi Lower: 21.55, Upper: 26.99
Estimates of CVg 36.6
Calculated Confidence Intervals of CVg Lower: 0.00, Upper: 45.73
Analytical CV 5.1
Calculated Confidence Intervals of Analytical CV Lower: 3.94, Upper: 7.22
Measurand Mean 0.7
Measurand Min
Measurand Max
Measurand SD

Coding

Parameter Score
Healthy Population true
Adult Population true
Weekly Samplings true
Reviewed true
Published true
Gender for Meta-Analysis Males
Snomed CT
NPU
LOINC
Comment
© 2018 - 2019 This website and its content is copyright of EFLM. All rights reserved. Any redistribution or reproduction of part or all of the contents in any form is prohibited other than the following: you may print or download to a local hard disk extracts for your personal and non-commercial use only you may copy the content to individual third parties for their personal use, but only if you acknowledge the website as the source of the material You may not, except with our express written permission, distribute or commercially exploit the content. Nor may you transmit it or store it in any other website or other form of electronic retrieval system.