| Parameter | Score |
|---|---|
| Matrix | Whole Blood EDTA |
| Measurand | Red cell distribution wide standard deviation (RDW-SD) |
| Unit | % / % |
| Overall Analytical Principle Method | Flow cytometry - Laser |
| Number of Subjects | 24 |
| Number of Subjects included in BV estimation | 17 |
| Ethinicity | Unknown |
| Age Mean / Median | 72.0 |
| Age Min | 58.0 |
| Age Max | 82.0 |
| Age SD | |
| Number of Males | 19 |
| Number of Females | 5 |
| State of Well Being |
Healthy (excl pregnant)
|
| Diseases | |
| Study Duration Length | 25 |
| Study Duration Units | Hour |
| Sampling Intervals, e.g. X sample(s) per Interval Unit | 1 |
| Interval Units | Hour |
| Sampling Times | Start Time:
End Time: |
| Number of Samples | 26 |
| Average Number of Samples Used for BV Estimation | 26.0 |
| Average Number of Replicates | 2.0 |
| Estimates of CVi | 0.37 |
| Calculated Confidence Intervals of CVi | Lower: 0.32, Upper: 0.42 |
| Estimates of CVg | 4.2 |
| Calculated Confidence Intervals of CVg | Lower: 3.15, Upper: 6.30 |
| Analytical CV | 0.47 |
| Calculated Confidence Intervals of Analytical CV | Lower: 0.44, Upper: 0.50 |
| Measurand Mean | 13.2 |
| Measurand Min | |
| Measurand Max | |
| Measurand SD |